                                        Min. working temperature
-90 °C
 
                                        Max. working temperature
220 °C
 
                                        Temperature stability
0.05 ± K
| Temperature | Pump level | Tempering medium | Cooling capacity 50 Hz | Cooling capacity 60 Hz | 
|---|---|---|---|---|
| 200 °C | 8 | Thermal oil | 4.5 kW | 4.5 kW | 
| 100 °C | 8 | Thermal oil | 4.5 kW | 4.5 kW | 
| 20 °C | 8 | Ethanol | 4.5 kW | 4.5 kW | 
| 10 °C | 8 | Ethanol | 4.45 kW | 4.45 kW | 
| 0 °C | 8 | Ethanol | 4.4 kW | 4.4 kW | 
| -10 °C | 8 | Ethanol | 4.4 kW | 4.4 kW | 
| -20 °C | 4 | Ethanol | 4.6 kW | 4.6 kW | 
| -30 °C | 4 | Ethanol | 4.6 kW | 4.6 kW | 
| -40 °C | 4 | Ethanol | 4.5 kW | 4.5 kW | 
| -50 °C | 4 | Ethanol | 4.2 kW | 4.2 kW | 
| -60 °C | 4 | Ethanol | 2.7 kW | 2.7 kW | 
| -70 °C | 4 | Ethanol | 1.4 kW | 1.4 kW | 
| -80 °C | 4 | Ethanol | 0.6 kW | 0.6 kW | 
| -90 °C | 2 | Ethanol | 0.2 kW | 0.2 kW | 
| Device | Device category | Document type | Language | Format | File size | |
|---|---|---|---|---|---|---|
| Chapter Circulation and process thermostats Overall brochure | Process thermostats | Brochure | 
                                    English
                                 | 4,84 MB | ||
| Integral XT New Generation | Process thermostats | Brochure | 
                                    English
                                 | 2,22 MB | ||
| LAUDA Command App | Process thermostats | Software | 
                                    English
                                 | EXE | 65,02 MB | |
| LAUDA Command App | Process thermostats | Brochure | 
                                    English
                                 | 1,43 MB | ||
| Integral XT New Generation with Flow Control Unit F-Gas | Process thermostats | Operating instruction | 
                                    English
                                 | 2,88 MB | ||
| Operating unit Command Touch | Accessories | Operating instruction | 
                                    English
                                 | 1,75 MB | ||
| Filling and draining unit FD 50 | Accessories | Operating instruction | 
                                    English
                                 | 4,46 MB | ||
| Assembly Filling and flow control system (based on FD 50 and MID 80) | Accessories | Operating instruction | 
                                    English
                                 | 1,24 MB | ||
| Automated refill FD 50 | Accessories | Operating instruction | 
                                    English
                                 | 1,09 MB | ||
| Integral XT New Generation F-Gas | Process thermostats | Operating instruction | 
                                    English
                                 | 14,17 MB | ||
| Integral XT New Generation CO2 | Process thermostats | Operating instruction | 
                                    English
                                 | 8,12 MB | ||
| Flow controller MID 80 | - | Operating instruction | 
                                    English
                                 | 2,16 MB | ||
| Integral XT New Generation Natural refrigerant | Process thermostats | Operating instruction | 
                                    English
                                 | 9,24 MB | 
There are already three products on your comparison list.
 To compare other products, please delete one of the products on your comparison list.