Operate your Integral process thermostat simply via a modern TFT display directly on the device or optionally via an intuitive operating unit with touch display and extended range of functions.
Working temperature min.
-45 °C
Max. working temperature
220 °C
Temperature stability
0.05 ± K
| Temperature | Pump level | Tempering medium | Cooling capacity 50 Hz |
|---|---|---|---|
| 200 °C | 8 | Thermal oil | 2.2 kW |
| 100 °C | 8 | Thermal oil | 2.2 kW |
| 20 °C | 8 | Ethanol | 2.1 kW |
| 10 °C | 8 | Ethanol | 2 kW |
| 0 °C | 8 | Ethanol | 1.8 kW |
| -10 °C | 8 | Ethanol | 1.4 kW |
| -20 °C | 4 | Ethanol | 1 kW |
| -30 °C | 4 | Ethanol | 0.55 kW |
| -40 °C | 4 | Ethanol | 0.2 kW |
| -45 °C | 2 | Ethanol | 0.05 kW |
| Device | Device category | Document type | Language | Format | File size | |
|---|---|---|---|---|---|---|
| Montage Befüll- und Durchflussregelsystem (basierend auf FD 50 und FC 80 MID) | Accessories | Operating instruction |
German
| 1,24 MB | ||
| LAUDA Command App | Process thermostats | Software |
German
| EXE | 95,34 MB | |
| LAUDA Command App | Process thermostats | Brochure |
German
| 1,43 MB | ||
|
Kapitel Umwälz- und Prozessthermostate Gesamtprospekt | Process thermostats | Brochure |
German
| 4,88 MB | ||
|
Integral XT New Generation mit Durchflussregler F-Gas | Process thermostats | Operating instruction |
German
| 2,91 MB | ||
|
Integral XT New Generation F-Gas | Process thermostats | Operating instruction |
German
| 15,04 MB | ||
|
Integral XT New Generation CO2 | Process thermostats | Operating instruction |
German
| 8,47 MB | ||
| Integral XT New Generation | Process thermostats | Brochure |
German
| 2,28 MB | ||
| Durchflussregler FC 80 MID | - | Operating instruction |
German
| 2,33 MB | ||
| Durchflussregler FC 80 C | - | Operating instruction |
German
| 2,52 MB | ||
| Befüll- und Entleereinheit FD 50 | Accessories | Operating instruction |
German
| 4,41 MB | ||
| Bedieneinheit Command Touch | Accessories | Operating instruction |
German
| 1,52 MB | ||
| Automatisierte Nachfüllung FD 50 | Accessories | Operating instruction |
German
| 1,09 MB | ||
|
Integral XT New Generation Natürliches Kältemittel | Process thermostats | Operating instruction |
German
| 10,05 MB |
There are already three products on your comparison list.
To compare other products, please delete one of the products on your comparison list.