Ubbelhode Viscometers for automatic cleaning

Viscosity measuring with Ubbelohde, Cannon-Fenske and Micro-Ostwald capillary viscometers

For automatic and stopwatch measurement. With screw connections and aspirating tube for permanent
installation. Recommended for automatic cleaning with LAUDA VRM modules.
• ISO 3105, DIN 51562, BS 188, NFT 60-100
• Filling volume: 18...22 ml
• Total length: approx. 290 mm
• Measurement accuracy: ±0.2 %
Also available in ASTM version

Your advantages at a glance

Your advantages at a glance

The glass capillary viscometer advantages Your benefits
Ubbelhode Viscometers for automatic cleaning
  • Ring marks with no detection disturbances
  • Precise, positioned, NIR permeable ring marks with exactly 40 mm clearance. The capillary constant is valid for stopwatch and automatic measuring systems.
Ubbelhode Viscometers for automatic cleaning
  • Corrosion resistant labelling
  • There is no risk of abrasion or etching off of the labeling even with the use of strong acids and aggressive solvents used as samples or cleaning agents.
Ubbelhode Viscometers for automatic cleaning
  • Ubbelohde for dilution series with calibration certificate and filter fit
  • Absolute measuring now also possible with calibrated dilution viscometers. The integrated frit ensures filtration when the sample dissolves in the viscometer.
  • Complete application from one supplier
  • LAUDA supplies all components for the viscosity application - clear-view thermostats, automatic viscosity measuring systems and glass capillary viscometers.
Ubbelhode Viscometers for automatic cleaning
  • Customer-specific variants available on request
  • LUADA supplies special variants as desired by the customer, specifically for adaption to automatic measuring systems.

Ubbelhode Viscometers for automatic cleaning

Type Cap. const. mm²/s² DIN/ASTM mm²/s PVS or iVisc mm²/s Ø i mm ±0,01 Cat. No. calibrated Cat. No. uncalibrated
0c 0,003 0,7...3 0,3…2 0,47 EGV 930 EGV 940
0a 0,005 1…5 0,5…3 0,53 EGV 931 EGV 941
I 0,01 2…10 0,7…7 0,63 EGV 932 EGV 942
Ic 0,03 6…30 2…20 0,84 EGV 933 EGV 943
II 0,1 20…100 6…60 1,13 EGV 934 EGV 944
IIc 0,3 60…300 20…200 1,50 EGV 935 EGV 945
III 1 200…1.000 60…600 2,01 EGV 936 EGV 946
IIIc 3 600…3.000 200…2.000 2,65 EGV 937 EGV 947
IV 10 2.000…10.000 600…6.000 3,60 EGV 938 EGV 948